Automatic Calibration
-
Ensure the Mini-OX is configured for automatic calibration. Refer to Setting the Calibration Type and Measurement Range
-
Place the sensor probe in the calibration gas, typically fresh air.
-
Allow the output tostabilize for at least 5 minutes (10 minutes if powering from cold).
-
Referring to page 9; Apply GND to the CALIBRATE input (pin 3) for a minimum 12s. During the 12s the CYCLE output (pin 4) and the green LED will go high/on, blink rapidly, go high/on, go low/off then return to cycling normally to indicate normal operation has resumed. At this point remove GND from pin 3.
NOTE:The output will now track to the correct value for the calibration gas. If calibrating in fresh air, the output value (RS232/Voltage/Current) given equates to 20.7% oxygen with an error of ± 0.2%. After calibration in fresh air the voltage output should read 8.28V. The current output should read 17.25mA, and the RS232 will stream a five character ASCII code representing the O2%.
-
Calibration is complete.
NOTE: New calibration values are stored in memory and retained on power loss.